SHIRAKAWA OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problems No Audible Alarm (1019); Display or Visual Feedback Problem (1184); Pressure Problem (3012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/08/2024
Event Type  malfunction  
Manufacturer Narrative
The device was returned and the evaluation found that there was light-emitting diode (led) failure on the front panel.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported, the insufflator continued to insufflate and would not alarm when the pressure became too high, the patient was insufflated too much.The issue occurred during an unspecified procedure.The procedure was completed.There were no reports of patient injury or harm.
 
Manufacturer Narrative
Additional findings noted the unit continues to insufflate when pressure is too high.Unable to confirm the initial report previously stated in section b5.Unit continues to insufflate and does not alarm when pressure is too high.Unable to confirm the initial report in section b5.Additional findings noted an led failure on the front panel due to the faulty front panel.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation and correction to h4.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device was returned to olympus and evaluated, with no confirmation of the issue.Based on the results of the investigation, the high insufflate with no alarm was not confirmed.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18810816
MDR Text Key336978978
Report Number3002808148-2024-01932
Device Sequence Number1
Product Code HIF
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/29/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2024
Was the Report Sent to FDA? No
Date Manufacturer Received03/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/16/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-