• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W.O.M. WORLD OF MEDICINE GMBH AIRSEAL IFS, 110V; INSUFFLATOR, LAPAROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W.O.M. WORLD OF MEDICINE GMBH AIRSEAL IFS, 110V; INSUFFLATOR, LAPAROSCOPIC Back to Search Results
Catalog Number AS-IFS1
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Air Embolism (1697)
Event Type  Injury  
Event Description
In the journal of medicine, st.Hamamatsu hospital, volume 23, issue 2 it was reported: a case of robot-assisted partial nephrectomy with gas embolis during surgery.It was reported that, ¿a case of robot-assisted partial nephrectomy in which gas embolism occurred during surgery¿.There was no report of the as-ifs1 having a malfunction.This report is being raised due to the reported injury of gas embolism.
 
Manufacturer Narrative
The device will not be returned, and no photographic evidence was provided.Therefore, a device malfunction cannot be verified.The service history review cannot be conducted as a serial number was not provided.A device history record (dhr) review cannot be conducted as a serial number was not provided.A two-year review of complaint history revealed there has been a total of 12 reports, regarding 13 devices, for this device family and failure mode.During this same time frame 1,224,402 devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be 0.00001.Per the instructions for use, the user is advised the following: improper placement of the insufflation instrument could cause gas penetrating a vessel or an internal organ, resulting in gas embolisms.To reduce the risk, use a low flow rate for the first insufflation and ensure that the insufflation instrument is correctly positioned.Check the position of the insufflation instrument immediately if the actual pressure rapidly reaches the nominal pressure value.Gas embolisms can also be caused by a high intra-abdominal pressure.Avoid high-pressure settings and close damaged blood vessels at once.We will continue to monitor for trends through the complaint system to assure patient safety.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIRSEAL IFS, 110V
Type of Device
INSUFFLATOR, LAPAROSCOPIC
Manufacturer (Section D)
W.O.M. WORLD OF MEDICINE GMBH
salzufer 8
berlin 10587
GM  10587
Manufacturer (Section G)
W.O.M. WORLD OF MEDICINE GMBH
salzufer 8
berlin 10587
GM   10587
Manufacturer Contact
robin drum
11311 concept blvd
largo, FL 33773
8653881978
MDR Report Key18810906
MDR Text Key336615705
Report Number1320894-2024-00048
Device Sequence Number1
Product Code HIF
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K190303
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberAS-IFS1
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/07/2024
Initial Date FDA Received02/29/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient EthnicityNon Hispanic
-
-