Catalog Number IAP-0700 |
Device Problem
No Device Output (1435)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/08/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4) n/a other remarks: n/a corrected data: n/a.
|
|
Event Description
|
It was reported that "when they were making their corridor transfer and unplugged the pump from the ac main power the fos ap disappeared completely.Once the pump was plugged back into ac power, the ap tracing was restored and the pump continued to work fine".No patient harm or injury.
|
|
Event Description
|
It was reported that "when they were making their corridor transfer and unplugged the pump from the ac main power the fos ap disappeared completely.Once the pump was plugged back into ac power, the ap tracing was restored and the pump continued to work fine".No patient harm or injury.
|
|
Manufacturer Narrative
|
(b)(4), the reported complaint of fos signal loss is not able to be confirmed.No part or recorder strip was returned to teleflex chelmsford for investigation.According to the expedited quality review (eqr) report, the fos board was replaced.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.No further action required at this time.The reported complaint will be monitored for any developing trends.
|
|