• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC AC3 OPTIMUS IABP NA/EMEA; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC AC3 OPTIMUS IABP NA/EMEA; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Catalog Number IAP-0700
Device Problem No Device Output (1435)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/08/2024
Event Type  malfunction  
Manufacturer Narrative
(b)(4) n/a other remarks: n/a corrected data: n/a.
 
Event Description
It was reported that "when they were making their corridor transfer and unplugged the pump from the ac main power the fos ap disappeared completely.Once the pump was plugged back into ac power, the ap tracing was restored and the pump continued to work fine".No patient harm or injury.
 
Event Description
It was reported that "when they were making their corridor transfer and unplugged the pump from the ac main power the fos ap disappeared completely.Once the pump was plugged back into ac power, the ap tracing was restored and the pump continued to work fine".No patient harm or injury.
 
Manufacturer Narrative
(b)(4), the reported complaint of fos signal loss is not able to be confirmed.No part or recorder strip was returned to teleflex chelmsford for investigation.According to the expedited quality review (eqr) report, the fos board was replaced.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.No further action required at this time.The reported complaint will be monitored for any developing trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AC3 OPTIMUS IABP NA/EMEA
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key18811075
MDR Text Key336757694
Report Number3010532612-2024-00139
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K162820
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberIAP-0700
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/08/2024
Initial Date FDA Received02/29/2024
Supplement Dates Manufacturer Received03/29/2024
Supplement Dates FDA Received04/01/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-