Model Number LF1937
Device Problems Device Alarm System (1012); Energy Output Problem (1431); Difficult to Open or Close (2921)
Patient Problem Hemorrhage/Bleeding (1888)
Event Date 02/01/2024
Event Type  malfunction  
Event Description
According to the reporter, during a procedure, the handpiece had start-up sound and the end sound rung, but the sealing was inadequa te/partial.There was possibility of insufficient squeezing of the grip.Bleeding was observed directly from the device seal.
 
Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the sealing was inadequate/partial and there was possibility of insufficient squeezing of the grip.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a gastric cancer surgery, the handpiece had start-up sound and the end sound rung, but the sealing was inadequate/partial.There was possibility of insufficient squeezing of the grip.There was a regrasp alert.Jaws of the ligasure were cleaned after every use.Procedure was completed using the reported device since there was a timing of completing the seal.Another handpiece worked fine with the same generator.Bleeding was observed directly from the ligasure seal but blood transfusion was not necessary.Patient's progress status is good.
 
Manufacturer Narrative
(rfr/fdd: alarm activation) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGASURE
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer (Section G)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18812572
MDR Text Key337546727
Report Number1717344-2024-00575
Device Sequence Number1
Product Code GEI
UDI-Device Identifier10884521705913
UDI-Public10884521705913
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K170869
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLF1937
Device Catalogue NumberLF1937
Device Lot Number32210260X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2024
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/05/2024
Initial Date FDA Received02/29/2024
Supplement Dates Manufacturer Received03/01/2024
Supplement Dates FDA Received03/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/09/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-