Model Number 133650 |
Device Problem
Mechanics Altered (2984)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 02/07/2024 |
Event Type
malfunction
|
Event Description
|
According to the reporter, during emergency (cabg) coronary artery bypass grafting, an open procedure, when clipping the vessel, two different clip appliers misfired on the same case.A small clip application was used to resolve the issue.
|
|
Manufacturer Narrative
|
D10 concomitant products: 133650, 133650 premium surgiclip iii 9.0 (lot#:p3j0844) medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the instrument was first test fired once in air so that the clip formation could be observed.The remaining clips were fired properly on test media with proper formation.The instrument was binding and the ratchet system was not functioning properly.The instrument was disassembled to reveal damage to the ratchet system.It was reported that the clips not loading properly.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur if the user attempts to either pull apart the handles prior to completing the handle compression or attempting to squeeze the handles prior to fully releasing the handles after completing a handle compression.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: the device contains a ratchet mechanism to ensure that a clip is not allowed to release until a full handle squeeze is applied.Ensure that the handles are squeezed together firmly as far as they will go.Failure to squeeze then handles completely can result in clip malformation and possible bleeding or leakage.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|
|
|