It was reported that a visions pv.018 catheter was used in a peripheral therapeutic procedure in the distal sfa.While passing through the lesion, the image disappeared.No patient injury reported.During the product return evaluation, missing material from the expanded single lumen was observed.This product problem is being submitted due to the missing material.There is a potential for harm if the malfunction were to recur.
|
This complaint was reviewed and investigated according to the manufacturer¿s policy.Blocks a2-a5: no information available.Blocks b6 & b7: no information available.Block c: not applicable for this device.Blocks d6 & d7: not applicable for this device.Blocks e1 & g2: country is japan.Block h6: the probable cause of the missing material is damage during use/handling as evidence by the sanguineous material observed within the lifted and missing expanded single lumen portion.Manipulation, strain, impact, and forces associated with use can affect the integrity of the device.Blocks h7 & h9: do not apply to this submission.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|