This device is classified as import for export, therefore 510k is not applicable.Model eg29-i10c-us is available in the usa with a 510k number k190805.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the pcb for cmos drive fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the pcb for cmos drive.In addition, our technician confirmed that the remote control buttons perforated, the insertion flexible tube buckled, the suction channel buckled, the r/l pulley wire stretched, and the distal body worn out; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|