Catalog Number 5517F302
Device Problem Unstable (1667)
Patient Problems Foreign Body Reaction (1868); Pain (1994)
Event Date 02/05/2024
Event Type  Injury  
Event Description
It was reported that the patient's right knee was revised due to global instability and pain.Intra-operatively, the following were observed: the locking wire of the insert broke, and the insert dislodged from the baseplate.The femoral component was articulating posteriorly, causing insert wear and metallosis due to articulating with the baseplate.The baseplate's posterior medial portion was broken off.Patient's activity level was normal, no trauma reported.The patient's cr knee (including patellar component) was converted to a ts knee.Rep confirmed that no further information will be released by the hospital or surgeon.
 
Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON P/A CR BEADED #3R
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
EI   NA
Manufacturer Contact
marisol santiago
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18814726
MDR Text Key336617012
Report Number0002249697-2024-00321
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200782
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 03/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/31/2021
Device Catalogue Number5517F302
Device Lot NumberBJR4N
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/19/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age73 YR
Patient SexFemale
Patient Weight96 KG
-
-