STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON P/A CR BEADED #3R; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
|
Back to Search Results |
|
Catalog Number 5517F302 |
Device Problem
Unstable (1667)
|
Patient Problems
Foreign Body Reaction (1868); Pain (1994)
|
Event Date 02/05/2024 |
Event Type
Injury
|
Event Description
|
It was reported that the patient's right knee was revised due to global instability and pain.Intra-operatively, the following were observed: the locking wire of the insert broke, and the insert dislodged from the baseplate.The femoral component was articulating posteriorly, causing insert wear and metallosis due to articulating with the baseplate.The baseplate's posterior medial portion was broken off.Patient's activity level was normal, no trauma reported.The patient's cr knee (including patellar component) was converted to a ts knee.Rep confirmed that no further information will be released by the hospital or surgeon.
|
|
Manufacturer Narrative
|
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
|
|