MEDTRONIC SINGAPORE OPERATIONS ATTESTA L DR MRI SURESCAN; PULSE GENERATOR, PERMANENT, IMPLANTABLE
|
Back to Search Results |
|
Model Number ATDRL1 |
Device Problems
Misconnection (1399); Impedance Problem (2950); High Capture Threshold (3266)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/19/2024 |
Event Type
Injury
|
Event Description
|
It was reported that post-operatively following implant the right ventricular (ra) lead exhibited elevated thresholds with a drop in impedance.Trouble shooting was performed by switching to unipolar configuration and showed that rv stimulation at high outputs resulted in stimulation of the atrium.It was also reported that the issues were due to a connection problem where in the rv lead was not fully inserted into the header of the implantable pulse generator (ipg).The lead and ipg remains in use. no patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Manufacturer Narrative
|
Corrected: b5.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that post-operatively following implant the right ventricular (rv) lead exhibited elevated thresholds with a drop in impedance.Trouble shooting was performed by switching to unipolar configuration and showed that rv stimulation at high outputs resulted in stimulation of the atrium.It was also reported that the issues were due to a connection problem where in the rv lead was not fully inserted into the header of the implantable pulse generator (ipg).The rv lead and ipg remains in use. no patient complications have been reported as a result of this event.It was further reported that the rv lead was re-inserted into the header and all the issues were resolved.
|
|