• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 PUREWICK FEMALE EXTERNAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 PUREWICK FEMALE EXTERNAL CATHETER Back to Search Results
Catalog Number PWFX30
Device Problem Patient Device Interaction Problem (4001)
Patient Problem Skin Inflammation/ Irritation (4545)
Event Date 02/07/2024
Event Type  Injury  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event Description
It was reported that the patient experienced irritation from the purewick female external catheter.Patient was going to stop using the product and to use a cream that their doctor had prescribed.
 
Event Description
It was reported that the patient experienced irritation from the purewick female external catheter.Patient was going to stop using the product and to use a cream that their doctor had prescribed.
 
Manufacturer Narrative
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be "inadequate material selection - materials of construction are not biocompatible or material surface is rough, abrasive or uncomfortable ".The dhr review could not be performed without a lot number.The instructions for use were found adequate and state the following: "instructions for use setup: 1.Connect the canister to wall suction and set to a minimum of 40mmhg continuous suction.Always use the minimum amount of suction necessary.If using the purewick¿ urine collection system, connect the canister to the unit and turn the unit on.Please consult the purewick¿ urine collection system user guide for further information.2.Using standard suction tubing, connect the purewick¿ female external catheter to the collection canister.Peri-care and placement: 3.Perform perineal care and assess skin integrity (document per hospital protocol).Separate legs, gluteus muscles, and labia.Palpate pubic bone as anatomical marker.4.With soft gauze side facing patient, align distal end of the purewick¿ female external catheter at gluteal cleft.Gently tuck soft gauze side between separated gluteus and labia.Ensure that the top of the gauze is aligned with the pubic bone.Slowly place legs back together once the purewick¿ female external catheter is positioned.Note: patient can be positioned on back, side lying, frog legged, or lying on back with knees bent and thighs apart (lithotomy position) prior to device placement.Removal: 5.To remove the purewick¿ female external catheter, fully separate the legs, gluteus, and labia.To avoid potential skin injury upon removal, gently pull the purewick¿ female external catheter directly outward.Ensure suction is maintained while removing the purewick¿ female external catheter.After use, this product may be a potential biohazard.Dispose of in accordance with applicable local, state and federal laws and regulations.Maintenance: 6.Replace the purewick¿ female external catheter at least every 8-12 hours or if soiled with feces or blood.Always assess skin for compromise and perform perineal care prior to placement of a new purewick¿ female external catheter." correction: e h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUREWICK FEMALE EXTERNAL CATHETER
Type of Device
FEMALE EXTERNAL CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key18815609
MDR Text Key336633767
Report Number1018233-2024-01077
Device Sequence Number1
Product Code NZU
UDI-Device Identifier00801741189050
UDI-Public(01)00801741189050
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Catalogue NumberPWFX30
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/07/2024
Initial Date FDA Received03/01/2024
Supplement Dates Manufacturer Received04/22/2024
Supplement Dates FDA Received04/24/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-