Model Number 459888 |
Device Problems
Low impedance (2285); Impedance Problem (2950); High Capture Threshold (3266)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/07/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Continuation of d10: 5076-45 lead implanted: on (b)(6) 2016 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during use of the left ventricular (lv) lead, a impedance lead integrity alert (lia) triggered, the lead exhibited impedance spikes, and drops, and high thresholds. the lv lead remains in use. it was also reported that the cardiac resynchronization therapy defibrillator (crt-d) exhibited drop in remaining longevity. the crt-d remains in use. no patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated the impedance of the left ventricular pacing lead had a spike.Analysis of the device memory indicated the impedance of the left ventricular pacing lead was below the expected lower range.Analysis of the device memory indicated pacing capture threshold in the left ventricle was high.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|
|