The hospital reported that during an endoscopic vessel harvesting procedure, vasoview hemopro 2 metal on tool split apart; the metal heater wire lifted.The procedure was completed using the same device.The (b)(6) further noted, "metal for the electrocautery began peeling off of device.It continued to work properly.I had never seen that happen before." there was no procedural delay and no patient harm.
|
Trackwise#:(b)(4).Updated sections: b4, g3, g6, h2, h3, h6, h10.The device was returned to the factory for evaluation on 02/14/2024.An investigation was conducted on 02/19/2024.A visual inspection was conducted.Signs of clinical use and evidence of blood was observed.Charred material was observed on the intact hot jaw.The heater wire was slightly flexed away from the hot jaw at the center of the hot jaw but remained attached to the base and tip of the hot jaw due to clinical use.There were no visual defects observed on the clear intact silicone insulation on both the cold and hot jaws.Based on the returned condition of the device as well as the evaluation results, the reported failure "material twisted/bent wire" was not confirmed.The lot # 3000360579 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
|