Model Number VH-4000
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/05/2024
Event Type  malfunction  
Manufacturer Narrative
Tw id (b)(4) the device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.
 
Event Description
The hospital reported that during an endoscopic vessel harvesting procedure, vasoview hemopro 2 metal on tool split apart; the metal heater wire lifted.The procedure was completed using the same device.The (b)(6) further noted, "metal for the electrocautery began peeling off of device.It continued to work properly.I had never seen that happen before." there was no procedural delay and no patient harm.
 
Manufacturer Narrative
Trackwise#:(b)(4).Updated sections: b4, g3, g6, h2, h3, h6, h10.The device was returned to the factory for evaluation on 02/14/2024.An investigation was conducted on 02/19/2024.A visual inspection was conducted.Signs of clinical use and evidence of blood was observed.Charred material was observed on the intact hot jaw.The heater wire was slightly flexed away from the hot jaw at the center of the hot jaw but remained attached to the base and tip of the hot jaw due to clinical use.There were no visual defects observed on the clear intact silicone insulation on both the cold and hot jaws.Based on the returned condition of the device as well as the evaluation results, the reported failure "material twisted/bent wire" was not confirmed.The lot # 3000360579 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEW HEMOPRO 2
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key18816533
MDR Text Key337202122
Report Number2242352-2024-00155
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00607567700406
UDI-Public00607567700406
Combination Product (y/n)N
PMA/PMN Number
K101274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVH-4000
Device Catalogue NumberVH-4000
Device Lot Number3000360579
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/05/2024
Initial Date FDA Received03/01/2024
Supplement Dates Manufacturer Received03/25/2024
Supplement Dates FDA Received03/25/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/11/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNKNOWN.
Patient Age79 YR
Patient SexFemale
Patient Weight91 KG
-
-