Catalog Number IAB-06840-U |
Device Problems
Leak/Splash (1354); Material Integrity Problem (2978)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/08/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that a leak was found in the catheter and inflation was not happening while connecting with machine.Another catheter was used in the same insertion site to complete the procedure.No patient injury or consequence reported.The patient's current condition is reported as "discharged".
|
|
Manufacturer Narrative
|
(b)(4).
|
|
Manufacturer Narrative
|
Qn# (b)(4).The reported complaint for iab "catheter found leak and inflation was not happening while connecting with machine" was not able to be confirmed as the product was not returned for investigation.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.No further action required at this time.Teleflex will continue to monitor and trend for reports of this nature.
|
|
Event Description
|
It was reported that a leak was found in the catheter and inflation was not happening while connecting with machine.Another catheter was used in the same insertion site to complete the procedure.No patient injury or consequence reported.The patient's current condition is reported as "discharged".
|
|