This device is classified as import for export, therefore 510k is not applicable.Model eg-2990k is available in the usa with a 510k number k131902.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd drive pcb fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ccd drive pcb.In addition, our technician confirmed that the control body fluid damage, the u/d pulley wire fluid damage, the up pulley wire fluid damage, the left pulley wire fluid damage, the right pulley wire fluid damage, the electrical pin connector fluid damage, the lg connector fluid damage, the lg connector fluid damage, and the light guide cable buckled; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|