The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd drive pcb fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ccd drive pcb.In addition, our technician confirmed that the electrical pin connector fluid damage, the forward body frame broken, the lcb (light carrying bundle) broken, the insertion flexible tube buckled, the suction channel buckled, the lg connector corroded, the remote control buttons leak, the root brace rubber (insertion flexible tube) cut, the body cover grip scratched, the distal body worn out, and the brand plate worn out; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|