• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN SURGICAL, S.A. MONOSYN VIOLET 2/0 (3) 70CM DS24 (M) RCP; SYNTHETIC ABSORBABLE MONOFILAR SUTURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN SURGICAL, S.A. MONOSYN VIOLET 2/0 (3) 70CM DS24 (M) RCP; SYNTHETIC ABSORBABLE MONOFILAR SUTURE Back to Search Results
Model Number C2022216
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Type  malfunction  
Manufacturer Narrative
G4: reported device not marketed in the u.S., however, similar devices or devices that share components, raw materials, process methods or other technological characteristics are registered within the u.S.Under pma/510(k) number: k011375.If additional information becomes available a follow up report will be submitted.
 
Event Description
It was reported an issue with monosyn suture.The client (veterinarian) reported that the thread was not attached to the needle when it was opened and it has happened (after asking other vets) that the thread has come loose during surgery, coming off the needle.
 
Manufacturer Narrative
Summary of investigation: there are no previous complaints of this code batch of which we manufactured and distributed in the market (b)(4) units.There are no units in stock in b.Braun surgical's warehouse.We have received an open and unused sample with the needle detached from the thread (thread is still wound on the pack).However, without closed samples a proper analysis cannot be performed.Considering that no other customer complaints have been received concerning this issue for this code-batch, we consider that this is an isolated unit, but the whole batch is correct.Batch manufacturing record: reviewed the batch manufacturing record, this product had no incidences related to this issue and was released fulfilling usp/ep and b.Braun surgical specifications.Conclusion root cause analysis: the root cause cannot be determined as no closed samples have been received and only one open sample received.Final conclusion: in spite of receiving a defective sample, without closed samples a suitable analysis cannot be performed.Nevertheless, we take note of this incidence and if any closed sample is received in the future, we will re-open the case and analyse it.Corrective measures: actions on distributed product of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONOSYN VIOLET 2/0 (3) 70CM DS24 (M) RCP
Type of Device
SYNTHETIC ABSORBABLE MONOFILAR SUTURE
Manufacturer (Section D)
B. BRAUN SURGICAL, S.A.
carretera de terrassa 121
rubí, barcelona 08191
SP  08191
Manufacturer (Section G)
B. BRAUN SURGICAL, S.A.
carretera de terrassa 121
rubí, barcelona 08191
SP   08191
Manufacturer Contact
martina laporte
carretera de terrassa 121
rubí, barcelona 08191
SP   08191
MDR Report Key18825192
MDR Text Key336753336
Report Number3003639970-2024-00068
Device Sequence Number1
Product Code GAM
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
SEE H11
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC2022216
Device Catalogue NumberC2022216
Device Lot Number13298M
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/05/2024
Initial Date FDA Received03/04/2024
Supplement Dates Manufacturer Received04/04/2024
Supplement Dates FDA Received04/05/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/20/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-