B. BRAUN SURGICAL, S.A. MONOSYN VIOLET 2/0 (3) 70CM DS24 (M) RCP; SYNTHETIC ABSORBABLE MONOFILAR SUTURE
|
Back to Search Results |
|
Model Number C2022216 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Type
malfunction
|
Manufacturer Narrative
|
G4: reported device not marketed in the u.S., however, similar devices or devices that share components, raw materials, process methods or other technological characteristics are registered within the u.S.Under pma/510(k) number: k011375.If additional information becomes available a follow up report will be submitted.
|
|
Event Description
|
It was reported an issue with monosyn suture.The client (veterinarian) reported that the thread was not attached to the needle when it was opened and it has happened (after asking other vets) that the thread has come loose during surgery, coming off the needle.
|
|
Manufacturer Narrative
|
Summary of investigation: there are no previous complaints of this code batch of which we manufactured and distributed in the market (b)(4) units.There are no units in stock in b.Braun surgical's warehouse.We have received an open and unused sample with the needle detached from the thread (thread is still wound on the pack).However, without closed samples a proper analysis cannot be performed.Considering that no other customer complaints have been received concerning this issue for this code-batch, we consider that this is an isolated unit, but the whole batch is correct.Batch manufacturing record: reviewed the batch manufacturing record, this product had no incidences related to this issue and was released fulfilling usp/ep and b.Braun surgical specifications.Conclusion root cause analysis: the root cause cannot be determined as no closed samples have been received and only one open sample received.Final conclusion: in spite of receiving a defective sample, without closed samples a suitable analysis cannot be performed.Nevertheless, we take note of this incidence and if any closed sample is received in the future, we will re-open the case and analyse it.Corrective measures: actions on distributed product of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
|
|