Model Number 6CFN |
Device Problem
Disconnection (1171)
|
Patient Problem
Unintended Extubation (4564)
|
Event Date 02/05/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
D10 concomitant product: 6cfn, 6cfn 6 fen trach cann w reus ic x1 (lot#23e0324jzx).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, the tracheostomy cannula's flange completely separated from the tube after three days of patient use.The tube was replaced.
|
|
Manufacturer Narrative
|
Additional information: d9, g3, h3 correction: h6 (codes rfr/ fdd) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Manufacturer Narrative
|
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the flange was separated from the tube and presented a damage apart from the pin breakage, the flange was missing the left pin, although the pin was received for evaluation.A dimensional test was performed on the tube and both are within the specifications.The depth of the holes were also measured, and both readings were within the specifications.The flange's pins were measured with a caliper and found to be within the specification.It was reported that the tracheostomy cannula's flange was completely separated from the tube after three days of patient use.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: the manufacturer recommends that the tracheostomy tube usage not exceed twenty-nine (29) days.Frequent and routine changes of the tracheostomy tube and accessories are recommended and should be evaluated by the attending physician.If you feel resistance when inserting the disposable inner cannula (dic) past the fenestration, do not force the inner cannula through the cannula.Notify your health care provider immediately.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|
|