• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROMODULATION IMPLANTABLE NEUROSTIMULATOR; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROMODULATION IMPLANTABLE NEUROSTIMULATOR; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS Back to Search Results
Model Number NEU_INS_STIMULATOR
Device Problems Human-Device Interface Problem (2949); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Stroke/CVA (1770); Hemorrhage/Bleeding (1888); Intracranial Hemorrhage (1891); Dysphasia (2195); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/25/2023
Event Type  Injury  
Event Description
It was reported that surgical procedure for implantable neurostimulator (ins) was not completed as patient (pt) was not able to complete surgery.Accessories were implanted, ins, lead, and extension were not.No symptoms reported.Additional information was received from the manufacturer representative (rep).It was an awake surgery, so they put the bhd and woke patient up for mer testing, he was responsive, they did some baseline testing and started mer, and some point along the way the patient was still awake but couldn't speak/answer questions, so they decided to abort.The cause was not determined, but they assumed there's a bleed along the trajectory.Patient has 2 bhd but nothing else, they didn't get to the leads part this all happened during mer testing.Patient will not be getting dbs anytime soon.
 
Manufacturer Narrative
Continuation of d10: product id b31000 lot# 082t26521 serial# implanted: (b)(6) 2023 product type accessory product id b32000 lot# 082m12423 serial# implanted: (b)(6) 2023 product type accessory section d information references the main component of the system.Other relevant device(s) are: product id: b31000, serial/lot #: (b)(6), ubd: 22-sep-2023, udi#: (b)(4); product id: b32000, serial/lot #: (b)(6), ubd: 04-may-2025, udi#: (b)(4)medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information received from the manufacturer¿s representative (rep) reported it was unclear what type of bleed the patient experienced as it was identified after the case was stopped during post-operative scans.It was believed the cause of the inability to speak/answer was due to a stroke, but this wasn¿t directly spoken by the physician.It was unclear how long the patient was unable to speak/answer for, but they were able to now and doing much better in their recovery.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE NEUROSTIMULATOR
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
Manufacturer (Section D)
MEDTRONIC NEUROMODULATION
7000 central ave ne
minneapolis MN 55432
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
7000 central ave ne
minneapolis MN 55432
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key18832614
MDR Text Key336858277
Report Number2182207-2024-01217
Device Sequence Number1
Product Code MHY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNEU_INS_STIMULATOR
Device Catalogue NumberNEU_INS_STIMULATOR
Was Device Available for Evaluation? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/04/2024
Supplement Dates Manufacturer Received03/04/2024
Supplement Dates FDA Received03/11/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-