It was reported that during a generator change with the new device no sensing was noted, and the pacing impedance went below the lower limit.The physician used a repair kit to repair the lead possibly due to some type of insulation breach.No further anomalies were observed, the device was functioning appropriately.One week post-procedure at in clinic check, noise was detected on a remote transmission.An attempt to reproduce the noise was done but it is unknown if the noise was reproducible.The patient will be monitored remotely.There were no adverse health consequences, the patient was stable.
|