It was reported that biclamp laparoscopic instrument was found to be broken during a laparoscopic gynecological operation.The accessory was used with an electrosurgical unit (esu).Specific information regarding the esu, settings, and other devices used were not provided.During the procedure, medical personnel noticed that the fastening screw in the joint area of the biclamp laparoscopic kelly insert was missing.Therefore, an x-ray was performed on the patient, but the screw was not found inside the patient.The screw also was not in the suction bag, on the instrument table, or on the floor.
|
The biclamp laparoscopic instrument was not returned for an evaluation [note: no issues were found upon the review of the lot's device history record (dhr)].However, a photograph was provided showing that the screw was missing from the instrument insert.Based upon similar reports involving this type of accessory, most likely, the instrument became damaged due to wear, mechanical stress (e.G., excessive levering or torsion), etc.However, no conclusive determination can be made at this time.Nevertheless, in the device's notes on use, it is stated that no excessive levering or exerting excessive forces on the instrument must be done, the product must be checked for damage before each use.Damaged/worn out instruments must not be used.Finally, if a root cause to the reported problem is determined upon an inspection of the instrument (if made available), etc.; a follow-up report will be filed.Erbe usa, inc.Is now closing the file on this event.
|