• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INOGEN X4 CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATORS (CRT-D)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INOGEN X4 CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATORS (CRT-D) Back to Search Results
Model Number G148
Device Problems Off-Label Use (1494); Low impedance (2285)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/22/2024
Event Type  malfunction  
Event Description
It was reported that the patient with this cardiac resynchronization therapy defibrillator (crt-d) device exhibited out of range low pacing impedance measurements, measuring less than 200 ohms on non-boston scientific left ventricular (lv) channel.The health care professional (hcp) had called in for magnetic resonance imaging (mri) programming assistance and noted this impedance measurements.Technical services (ts) stated that this could be a lead integrity issue.The hcp had a non-conditional form and boston scientific representative confirmed that mri was successfully performed.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN X4 CRT-D
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATORS (CRT-D)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18836484
MDR Text Key337017474
Report Number2124215-2024-13315
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534591
UDI-Public00802526534591
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/10/2016
Device Model NumberG148
Device Catalogue NumberG148
Device Lot Number105206
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/22/2024
Initial Date FDA Received03/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/12/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age85 YR
Patient SexFemale
-
-