Model Number 176625 |
Device Problems
Mechanical Problem (1384); Material Protrusion/Extrusion (2979)
|
Patient Problem
Unintended Radiation Exposure (4565)
|
Event Type
Injury
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during a laparoscopic cholecystectomy, at the cystic duct, the surgeon was able to squeeze the handle, and the surgeon complained that it felt the clip was tearing through the vessel and it felt even if there was a wider clip.The surgeon also noted that the sharpness between the clip and the proximal jaw was quite sharp.An endoscopic retrograde cholangiopancreatography (ercp) was carried out, a cystic duct leakage was discovered, and a stent was used.The surgical time was extended for 30 minutes.A competitor's device was used to resolve the issue.
|
|
Manufacturer Narrative
|
Additional information: b5, g3 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
According to the reporter, postoperatively, of a laparoscopic cholecystectomy, at the cystic duct, the surgeon was able to squeeze the handle, and the surgeon complained that it felt the clip was tearing through the vessel and it felt even if there was a wider clip.The surgeon also noted that the sharpness between the clip and the proximal jaw was quite sharp.An endoscopic retrograde cholangiopancreatography (ercp) was carried out, a bile leak from the cystic duct was discovered, and a stent was used.It was noted that as a result a patient treatment was needed.A competitor's device was used to resolve the issue.
|
|
Search Alerts/Recalls
|
|
|