• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. LEVEL 1 LOW FLOW - HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. LEVEL 1 LOW FLOW - HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problem No Display/Image (1183)
Patient Problem Insufficient Information (4580)
Event Date 02/01/2024
Event Type  malfunction  
Manufacturer Narrative
B3.Date of event: unknown.No information has been provided to date.D4.Serial number and h4.Manufacture date are unknown; no information has been provided to date.D5.Other operator of device: operator of device is unknown.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the lcd for temperature display was broken.There was unknown patient involvement and unknown patient harm/adverse event reported.
 
Manufacturer Narrative
Evaluation codes: updated.Device evaluation: one device was received.Per visual inspection, the reservoir cover was cracked, and the drain cap and rubber feet were missing.Functional testing confirmed that the liquid crystal display (lcd) of the main unit was damaged, and the temperature was not displayed.The root cause was damage to the main board.It was unknown what caused it to become damaged.The product's history records were reviewed and there were no non-conformance's nor service-related issues that would have resulted in the reported complaint.The main pc board assembly and water tank cover were replaced.Also, attached the rubber feet and drain cap that fell off.The device passed all functional tests after repair.D4 - unique identifier (udi) #, d4 - serial number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 LOW FLOW - HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane n
minneapolis, MN 55442
2247062300
MDR Report Key18837398
MDR Text Key336898032
Report Number3012307300-2024-01147
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier50695085820909
UDI-Public(01)50695085820909(17)190115(10)22007158
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberHL-90
Device Catalogue NumberHL-90JP
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer02/14/2024
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/06/2024
Initial Date FDA Received03/05/2024
Supplement Dates Manufacturer Received03/24/2024
Supplement Dates FDA Received04/18/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/15/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-