GLOBUS MEDICAL, INC. RESONATE; RESONATE ANTERIOR CERVICAL PLATE, 2-LEVEL, 32MM Back to Search Results
Model Number 1194.2032
Device Problem Break (1069)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Event Description
It was reported that a screw is backing out of the resonate plate post-operatively.
 
Manufacturer Narrative
This report is being resubmitted for an incident that occurred earlier.The device was not available for evaluation as it remains in the patient.The immediate post-operative image all screws appear to be fully seated.The follow-up images show the head of a bottom screw is partially protruding from the plate.It is unclear whether the cause was a damaged slider or if the screw was not fully seated in the plate.No determinations could be made as to the cause of the reported issue.The following sections are been updated: b4, e1, h2, h6, h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESONATE
Type of Device
RESONATE ANTERIOR CERVICAL PLATE, 2-LEVEL, 32MM
Manufacturer (Section D)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon PA 19403
Manufacturer (Section G)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon PA 19403
Manufacturer Contact
daniel paul
2560 general armistead avenue
audubon, PA 19403
6109301800
MDR Report Key18838074
MDR Text Key337155540
Report Number3004142400-2023-00138
Device Sequence Number1
Product Code KWQ
UDI-Device Identifier00193982154837
UDI-Public00193982154837
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K192314
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/05/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1194.2032
Was Device Available for Evaluation? No
Date Manufacturer Received08/09/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age48 YR
Patient SexFemale
Patient RaceBlack Or African American
-
-