• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) UNIFY ASSURA CRT-D RF HV; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) UNIFY ASSURA CRT-D RF HV; NO MATCH Back to Search Results
Model Number CD3357-40Q
Device Problems Pacing Problem (1439); Environmental Compatibility Problem (2929)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the implantable cardioverter defibrillator exhibited an inappropriate rate increase.Further information was requested but was unavailable at this time.
 
Event Description
New information notes the issue was discovered by a hospital device technician.Prior to the magnetic resonance imaging (mri) scan, the implant was programmed as ordered by the physician orders to voo mode without the presence of a field representative.The mri scan was conducted in an off-label mri environment.The issue resolved.No further action was required.The patient was stable before, during and after the incident.*(voo mode induces sequential asynchronous ventricular pacing, without inhibition by intrinsic events.).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY ASSURA CRT-D RF HV
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18839733
MDR Text Key336924401
Report Number2017865-2024-34154
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508216
UDI-Public05414734508216
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2023
Device Model NumberCD3357-40Q
Device Lot NumberA000115176
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/15/2024
Initial Date FDA Received03/05/2024
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/29/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/13/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA LEAD RIGHT VENTRICULAR LEAD; QUICK FLEX LEAD; TENDRIL ST OPTIM LEAD
Patient Age86 YR
Patient SexFemale
-
-