MAUDE Adverse Event Report: MEDTRONIC IRELAND MICRA; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND MICRA; INTRODUCER, CATHETER Back to Search Results
Model Number MI2355A
Device Problems Mechanical Problem (1384); Use of Device Problem (1670); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/22/2024
Event Type  Injury  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during implant of the leadless implantable pulse generator (ipg), recapture was attempted due to high thresholds.However , the alignment could not be achieved and the tether was cut repeatedly which caused the device to dislodge.The device was collected using a snare.It was also noted that the introducer used in the procedure was more than six month passed the use by date.The leadless ipg was attempted/not used and replaced. no patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the introducer was returned and analyzed.The distal seal of the delivery system introducer was torn.The analyst noted the introducer was returned with the dilator fully inserted in the sheath of the introducer.It is not known what the use by date of the introducer was as the packaging was not returned nor the lot number returned.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRA
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway,ie
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway,ie
EI  
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key18839744
MDR Text Key336924546
Report Number9612164-2024-01102
Device Sequence Number1
Product Code DYB
UDI-Device Identifier00643169359888
UDI-Public00643169359888
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K132030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/05/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/04/2021
Device Model NumberMI2355A
Device Catalogue NumberMI2355A
Device Lot NumberMI2355A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/21/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-