BOSTON SCIENTIFIC CORPORATION EPIC VASCULAR; STENT, ILIAC Back to Search Results
Model Number 39200
Device Problem Premature Activation (1484)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/14/2024
Event Type  malfunction  
Event Description
It was reported that stent inadvertent deployment occurred.A 10x60x75 epic vascular stent balloon was selected for use in an iliac angioplasty.The device was flushed and loaded onto the wire; however, it was noted that the distal portion of the stent was exposed and partially deployed.The procedure was completed with a new 10x40 epic stent.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIC VASCULAR
Type of Device
STENT, ILIAC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18840299
MDR Text Key336928555
Report Number2124215-2024-12957
Device Sequence Number1
Product Code NIO
UDI-Device Identifier08714729805052
UDI-Public08714729805052
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P110035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number39200
Device Catalogue Number39200
Device Lot Number0033036460
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/14/2024
Initial Date FDA Received03/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/12/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
Patient RaceWhite
-
-