BOSTON SCIENTIFIC CORPORATION MOMENTUM CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
|
Back to Search Results |
|
Model Number G124 |
Device Problems
Signal Artifact/Noise (1036); Premature Discharge of Battery (1057); Defective Device (2588); Battery Problem (2885)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/21/2024 |
Event Type
Injury
|
Event Description
|
It was reported that the battery of this cardiac resynchronization therapy defibrillator (crt-d) was suspected to be depleting prematurely and the left ventricular (lv) lead exhibited low out-of-range pace impedance measurements less than 200 ohms and noise.A request was made to have data from this device analyzed.Data analysis confirmed the battery power consumption was high and irregular, consistent with a latent pace block anomaly in the analog integrated circuit that can impact lead integrity.Additional information received indicated that this crt-d was explanted and replaced, and the lv was capped and replaced with the secondary previously implanted lv lead.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
|