• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VERATHON MEDICAL ULC GLIDESCOPE SPECTRUM SMART CABLE; LARYNGOSCOPE, RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VERATHON MEDICAL ULC GLIDESCOPE SPECTRUM SMART CABLE; LARYNGOSCOPE, RIGID Back to Search Results
Model Number 0800-0543
Device Problems Erratic or Intermittent Display (1182); Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/21/2024
Event Type  malfunction  
Manufacturer Narrative
The customer declined the option to have their glidescope spectrum smart cable returned to verathon for evaluation.Since the device was not returned to verathon for evaluation, the cause could not be determined; however, it is likely that the reported damage to the cable's hdmi connector pins may have caused or contributed to the image issues.The glidescope video laryngoscopes operations and maintenance manual (omm) notes that "before every use, ensure that the instrument is operating correctly and has no sign of damage.Do not use this product if the device appears damaged." verathon followed up with the customer and restated the importance of checking the device before its use in a procedure.Review of complaint history for the reported smart cable serial number (b)(6) did not identify any previous complaints.Trending analysis for glidescope spectrum smart cables does not identify any trends exceeding acceptable limits.Review of the system risk assessment confirmed that the risk associated with the hazardous scenario has been adequately captured and characterized.Corrective action is not required at this time.Verathon will continue to monitor for trends.
 
Event Description
A customer reported that during a patient procedure, using a glidescope spectrum smart cable, the image went blank as soon as they inserted the endotrachial tube.Furthermore, the customer mentioned that the hdmi connector pins of the cable appeared damaged.No delay in the procedure, use of a backup device, or harm to the patient was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE SPECTRUM SMART CABLE
Type of Device
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, V5C 5 A9
CA  V5C 5A9
Manufacturer (Section G)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, V5C 5 A9
CA   V5C 5A9
Manufacturer Contact
corey kasbohm
20001 n creek pkwy
bothell, WA 98011-8218
4256295760
MDR Report Key18840730
MDR Text Key336970352
Report Number9615393-2024-00042
Device Sequence Number1
Product Code CCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 02/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0800-0543
Device Catalogue Number0800-0544
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/21/2024
Initial Date FDA Received03/05/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/20/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-