• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 BD PREVUE II PERIPHERAL VASCULAR ACCESS SYSTEM 20MM TRADITIONAL PROBE; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 BD PREVUE II PERIPHERAL VASCULAR ACCESS SYSTEM 20MM TRADITIONAL PROBE; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Catalog Number 9770703
Device Problem Poor Quality Image (1408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/10/2024
Event Type  malfunction  
Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The device was returned to service facility for evaluation.During evaluation, the reported issue of probe has a dark area on the image was confirmed.The probe has 2 failed transducer elements which cause a dark spot in the ultrasound image.The root cause is due to an internal failure of the probe.H3 other text : evaluation summary findings in h:11.
 
Event Description
Observation found during evaluation at bd service center: the probe has 2 failed transducer elements which cause a dark spot in the ultrasound image.This issue is due to an internal failure of the probe.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PREVUE II PERIPHERAL VASCULAR ACCESS SYSTEM 20MM TRADITIONAL PROBE
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer Contact
shelly gilbert
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key18841535
MDR Text Key336968327
Report Number3006260740-2024-00921
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741214950
UDI-Public(01)00801741214950
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K211193
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number9770703
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/14/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/10/2024
Initial Date FDA Received03/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-