• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Catalog Number 1800350-38
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Angina (1710); Obstruction/Occlusion (2422); Prolapse (2475)
Event Date 01/31/2024
Event Type  Injury  
Manufacturer Narrative
D4: the udi is unknown as the part/lot number was not provided.Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Event Description
It was reported that a xience skpoint stent was implanted.Ten minutes after the case, stenosis occurred.Upon re-accessing the target site, plaque was noted to be protruding between cells.Additional intervention was performed as treatment.There was no adverse patient sequela.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.The reported patient effect of occlusion, prolapse and angina are listed in the xience skypoint everolimus eluting coronary stent systems instructions for use as a known patient effects of coronary stenting procedures.A conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined; however, the subsequent treatment appears to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling; therefore, no product-related corrective action will be implemented in this case.
 
Event Description
Subsequent to the initially filed reports, it was reported the patient presented with st elevation myocardial infarction (stemi).The procedure was performed to treat a lesion in the left anterior descending (lad) coronary artery with no calcification, mild tortuosity and 60% stenosis.Post stent implant, the patient experienced chest pain and intravascular ultrasound (ivus) confirmed plaque was protruding between cells.An unspecified drug coated balloon dilatation catheter was used to treat the plaque protrusion.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18842012
MDR Text Key336956443
Report Number2024168-2024-02769
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier08717648230257
UDI-Public(01)08717648230257(17)260611(10)3062241
Combination Product (y/n)Y
Reporter Country CodeSP
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number1800350-38
Device Lot Number3062241
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/12/2024
Initial Date FDA Received03/05/2024
Supplement Dates Manufacturer Received03/25/2024
Supplement Dates FDA Received03/28/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/12/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-