Model Number 101967-950
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/02/2023
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted retrospectively as part of internal review.After reviewing the sensor data in dms we can see that ec39 sensor replacement alert was triggered on 2 july 2023 at 12:53 pm.The sensor was therefore replaced and returned for investigation.Further analysis of the in vivo glucose data shows that the rate of change in the glucose values frequently exceeded 5mg/dl/min.This noise is the most probable cause of the failure.The investigation shows the system correctly disabled the sensor due to performance failure and the system's self-test functions are working normally.The failure mode could not be reproduced during the returned product analysis testing.This may occur in some instances where the failure mode that presents itself in the body is not reproduced in the lab.
 
Event Description
On 03 july 2023,senseonics was made aware of an incident where user received an early sensor replacement alert resulting in an early sensor removal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE SENSOR
Type of Device
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20876 7005
Manufacturer (Section G)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20876 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown, MD 20876-7005
MDR Report Key18842933
MDR Text Key337018180
Report Number3009862700-2024-00471
Device Sequence Number1
Product Code QHJ
UDI-Device Identifier00817491023308
UDI-Public817491023308
Combination Product (y/n)Y
Reporter Country CodeSP
PMA/PMN Number
P160048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 03/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/05/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/08/2023
Device Model Number101967-950
Device Catalogue NumberFG-4500-50-302
Device Lot NumberWP09007
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/05/2023
Date Manufacturer Received09/29/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/26/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-