It was reported that on february 9, a myosure procedure was performed and during the procedure, the physician dilated, and inserted the myosure xl scope, and the fluent deficit immediately jumped to 500ml¿s.The procedure was aborted.The physician took the patient back for laparoscopic surgery that evening and visualized where the scope or dilator went through.They sutured the perforation, then did the mysoure portion under ultrasound and lap guidance.No additional information available.
|
Device identifier: (b)(4).D4: lot and serial number of the device not provided by the complainant; therefore, the udi, expiration and manufacturing dates are not known.Device history record (dhr) review was unable to be conducted for the disposable device as the identification numbers were not provided by the complainant.H3: the device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported and a definitive root cause for the reported event could not be determined.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
|