MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
|
Back to Search Results |
|
Model Number 97715 |
Device Problems
Overheating of Device (1437); Delayed Charge Time (2586); Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
|
Patient Problems
Pain (1994); Burning Sensation (2146)
|
Event Date 02/24/2024 |
Event Type
malfunction
|
Event Description
|
Information was received from a patient regarding an external device.The reason for call was patient reports having issues with recharger.Pt states the recharger paddle got super hot and the relay box was also super hot.Pt mentioned trying to charge 2.5 hours and went from 30-40%, pt states he doesnt want to be in pain and needs to charge.An email was sent to the repair department to replace the device.Agent sent request to repair for rtm.Pt mentioned the device fried on friday night but quit saturday.Agent had a hard time gathering this information.
|
|
Manufacturer Narrative
|
Date is approximate.Month and year are confirmed valid.Section d references the main component of the system.Other medical products in use during the event include: brand name intellis; product id 97755-s (b)(6); product type: 0213-recharger.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Continuation of d10: product id 97755-s serial (b)(6): product type recharger medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Manufacturer Narrative
|
H3: analysis of the recharger, found a non-conformance.The recharge antenna failed, due to a "poor recharge quality" message.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank, because the information is currently, unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|