Related manufacturer report number: 2017865-2024-34242, related manufacturer report number: 2017865-2024-34243, related manufacturer report number: 2017865-2024-34248, related manufacturer report number: 2017865-2024-34251, related manufacturer report number: 2017865-2024-34254.It was reported during device upgrade that the right ventricular lead and atrial leads were stuck together, so both were explanted.A left ventricular lead was positioned but dislodged, as did its two replacements.During positioning attempts the newly implanted atrial lead was dislodged and had to be exchanged.A replacement right ventricular (rv) lead was positioned but dislodged.Attempt to reposition the rv lead was unsuccessful as the helix would not re-extend.The patient was stable.
|