• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS EXERA III DUODENOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS EXERA III DUODENOVIDEOSCOPE Back to Search Results
Model Number TJF-Q190V
Device Problems Crack (1135); Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/11/2024
Event Type  malfunction  
Event Description
It was observed that during the device evaluation, the duodenvideoscope adhesive between distal end and server plug-in (z-block) is cracked and has a gap.There were no reports of patient involvement.
 
Manufacturer Narrative
The device was returned and the evaluation found the following: distal end has residual liquid; adhesive between distal end and server plug-in (z-block) is cracked and has a gap; due to wear of forceps wire (k-wire), the forceps raising angle does not meet the standard value; indication on grip is unclear; air/water cylinder has no color; suction cylinder has no color; label on scope connector is damaged; distal end is shaved; and universal cord has coating peeling.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be determined.It is likely the adhesive between the distal end and server plug-in (z-block) became cracked and had a gap because of applied physical stress (such as hitting or dropping) and/or chemical stress (such as a chemical solution used).As a result of the adhesive failure, normal aeration could not remove water and residual liquid may have remained at the distal end.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III DUODENOVIDEOSCOPE
Type of Device
DUODENOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18846999
MDR Text Key337652550
Report Number9610595-2024-04711
Device Sequence Number1
Product Code FDT
UDI-Device Identifier04953170405563
UDI-Public04953170405563
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K220587
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTJF-Q190V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/19/2024
Initial Date FDA Received03/06/2024
Supplement Dates Manufacturer Received03/27/2024
Supplement Dates FDA Received04/09/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/29/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-