|
Catalog Number 05.001.201 |
Device Problems
Appropriate Term/Code Not Available (3191); Physical Resistance/Sticking (4012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2024 |
Event Type
malfunction
|
Event Description
|
It was reported from germany that during service and evaluation, it was determined that the battery handpiece device had the following failures: seized mechanics, component damaged, leak tightness test failure, sticky trigger, will not hold/secure battery and will not run.It was further determined that the device failed pretest on general condition, leakage test using bubble emission technique, check for mechanical free moving, check for sticky triggers, check falling out protection (steel ring) and check general function of device.It was noted in the service order that the device had an unspecified malfunction.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown but was noted to have occurred in january 2024.All available information has been disclosed.If additional information should become available, an additional report will be submitted accordingly.
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.H10 additional narrative: (b)(4) the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to improper maintenance.
|
|
Search Alerts/Recalls
|
|
|