MAQUET CARDIOVASCULAR LLC VASOVIEW HEMOPRO; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number VH-3500
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/15/2024
Event Type  malfunction  
Event Description
The hospital reported that during an endoscopic vein harvesting procedure, vasoviewhempro vh-3500 jaws not opening all the way.New device was gotten out to complete case.No delay in case time.No patient effects.
 
Manufacturer Narrative
Tw id# (b)(4).The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.
 
Event Description
N/a.
 
Manufacturer Narrative
Trackwise#: (b)(4).The device was returned to the factory for evaluation on 02/20/2024.An investigation was conducted on 02/20/2024.A visual inspection was conducted.Signs of clinical use and evidence of blood and charred material were observed.The heater wire and gray silicone insulation of the jaws were observed to be intact with no visual defects.A mechanical evaluation was conducted.The blue toggle was manipulated to open and close the jaws.The jaws opened fully when the toggle was slid backward.Based on the returned condition of the device and investigation results, the reported failure "mechanical problem; jaws will not open completely", was not confirmed.The lot # 3000363947 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEW HEMOPRO
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key18851742
MDR Text Key337091902
Report Number2242352-2024-00189
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00607567701250
UDI-Public00607567701250
Combination Product (y/n)N
PMA/PMN Number
K153194
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/06/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVH-3500
Device Catalogue NumberVH-3500
Device Lot Number3000363947
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/02/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/04/2024
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age61 YR
Patient Weight64 KG
-
-