Catalog Number 07P70-30
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/12/2024
Event Type  malfunction  
Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient information was included.Additional patient details are not available.
 
Event Description
The customer observed a falsely elevated alinity i free t4 result for one patient.The following data was provided: sample id (b)(6) initial result, on 15feb, was >5.00, repeat was 1.38 ng/dl.There was no impact to patient management reported.
 
Event Description
The customer observed a falsely elevated alinity i free t4 result for one patient.The following data was provided: sample id (b)(4) initial result, on (b)(6) was >5.00, repeat was 1.38 ng/dl.There was no impact to patient management reported.
 
Manufacturer Narrative
The complaint investigation for a falsely elevated alinity i free t4 results included a search for similar complaints, the review of complaint text, trending data, labeling, device history records, and testing of retained reagent kit.Return testing was not completed, as returns were not available.Accuracy testing was performed using panels, which mimics patient samples, and an in-house retained kit of lot 57607ud00, stored at the recommended storage condition.All specifications were met indicating that the lot is performing acceptably.Trending review determined no related trend for the issue for the product.Device history record review did not identify any non-conformances or deviations for the likely cause lot and complaint issue.Manufacturing documentation for the likely cause lot was reviewed and did not identify any issues.Labeling was reviewed and found to adequately address the issue under review.Based on the information provided and abbott diagnostics¿ complaint investigation, no systemic issue or deficiency of alinity i free t4, lot number 57607ud00, was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALINITY I FREE T4 REAGENT KIT
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ABBOTT IRELAND DIAGNOSTICS DIVISION
lisnamuck
co. longford
longford N39E9 32
EI  N39E932
Manufacturer (Section G)
ABBOTT IRELAND DIAGNOSTICS DIVISION
lisnamuck
co. longford
longford N39E9 32
EI   N39E932
Manufacturer Contact
nicole jenne
max-planck-ring 2
post market surveillance
wiesbaden 65205
GM   65205
6122582960
MDR Report Key18851743
MDR Text Key337109402
Report Number3005094123-2024-00095
Device Sequence Number1
Product Code CEC
UDI-Device Identifier00380740121822
UDI-Public(01)00380740121822(17)241031(10)57607UD00
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K173122
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number07P70-30
Device Lot Number57607UD00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/16/2024
Initial Date FDA Received03/06/2024
Supplement Dates Manufacturer Received04/11/2024
Supplement Dates FDA Received04/11/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/28/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ALNTY I PROCESSING MODU, 03R65-01, (B)(6). ; ALNTY I PROCESSING MODU, 03R65-01, (B)(6).
-
-