SHIRAKAWA OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problem Intermittent Loss of Power (4016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/16/2024
Event Type  malfunction  
Event Description
The customer reported to olympus the insufflation unit kept resetting itself during an unspecified procedure.The procedure outcome was not reported.There was no report of patient injury or medical intervention associated with this event.
 
Manufacturer Narrative
The device was returned to olympus and evaluated.The customer¿s allegation was not confirmed.The investigation is ongoing and follow-up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 10 years since the subject device was manufactured.Based on the results of the investigation, the reported issue (device resetting itself) was not reproduced.Therefore, the root cause could not be determined.This supplemental report includes additional information received from the customer.B5 updated accordingly.Also, a correction has been made to d4 and g2 to provide information that was inadvertently not included in the initial medwatch.Olympus will continue to monitor field performance for this device.
 
Event Description
It was reported that the incident occurred during the middle of a procedure to diagnose an ectopic pregnancy.There were no error messages displayed at the time of the event.It was confirmed that the device was inspected prior to use per the olympus instructions for use (ifu) manual.Although the incident resulted in a 5-minute delay, the procedure was completed using the same device.The patient did not require any additional medical intervention and was not impacted by the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18853242
MDR Text Key337845149
Report Number3002808148-2024-02115
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public4953170324147
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/06/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/19/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/22/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-