It was reported that on (b)(6) 2024, a 24mm amplatzer septal occluder was selected for an implant using a 10f amplatzer trevisio intravascular delivery system.At the beginning of the preparation of the device, it was not screwing well into the release cable, making it difficult to screw it properly.During the procedure, device had a cobra effect in the deployment of the left disc, device was recapture and remove from the patient.The device was checked in clean water and its cobra shape did not improve.No interaction with cardiac structures during deployment and no angulation or kink noticed in the delivery system.A new 22mm amplatzer septal occluder was selected and successfully completed the procedure.The patient is stable.
|
An event of device deformity could not be confirmed.The device was returned to abbott for investigation and the device met functional specifications when analyzed under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.Reportedly, it was indicated that there was no anatomical interference, there was no angulation or kink in the delivery system upon deployment, and an 10f delivery sheath was used.As recommended per the instructions for use for 24 mm amplatzer ¿ septal occluder the 9f was intended use which could have contributed to the reported event of deformity.There were no complaints associated with any other devices from the lot.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|