An event of device deformity was reported.A returned device inspection could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.There were no complaints associated with any other devices from the lot.Indication in the field indicated that 10f delivery system was used, there was no anatomical interference, and there was no angulation or kink in the delivery system.Based on the information received, the cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with regards to manufacture, design, or labeling.Na.
|
It was reported that on (b)(6) 2024, a 20mm amplatzer septal occluder was selected for an implant using a 10f trevisio delivery system.On deployment of the left atrial disc the device had a "cobra" appearance.This was withdrawn and deployed again but still had the same appearance.Device was removed from the patient prior to released from the cable.No interaction with cardiac structures during deployment and no angulation or kink noticed in the delivery system.A new 20 mm amplatzer septal occluder device was requested but not available, so a 19mm amplatzer septal occluder device was successfully deployed using the same delivery system.The patient is stable.
|