MICROVENTION, INC. WEB SL; INTRASACCULAR FLOW DISRUPTION DEVICE Back to Search Results
Model Number W5-4-2-MVI-3
Device Problem Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/28/2023
Event Type  malfunction  
Event Description
It was reported, web 4x2 was used in a w:3.2mm x h:2.7mm acomm aneurysm with tortuosity.Web device was positioned correctly and detachment was attempted.Controller was working with proper battery but no deflection.Attempt was made to advance via over proximal marker and device kept resheathing.Another attempt was made with wdc, then new wdc with same reference number.Another attempt was made to advance via but device did not detach and the device was resheathed and removed device.No patient injury or intervention was performed.
 
Manufacturer Narrative
The investigation of the returned web system found the heater coil windings stretched.The device failed continuity and resistance testing, which is consistent with the stretched heater coil windings.However, the heater coil pet and implant tether were found melted, indicating that the device was activated using a detachment controller during the procedure; therefore, the damage to the heater coil windings likely occurred post-activation.The stretched heater coil windings are an indication that the tether could have been caught in between the coil windings and caused the heater coil to stretch when the delivery system was pulled/retracted during the procedure.The returned detachment controller was found to function as intended and would not have caused or contributed to the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WEB SL
Type of Device
INTRASACCULAR FLOW DISRUPTION DEVICE
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise
aliso viejo, CA 92656
7142478000
MDR Report Key18862078
MDR Text Key337780490
Report Number2032493-2024-00200
Device Sequence Number1
Product Code OPR
UDI-Device Identifier00842429110812
UDI-Public(01)00842429110812(11)230725(17)260630(10)0000388654
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P170032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberW5-4-2-MVI-3
Device Lot Number0000388654
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/21/2024
Initial Date FDA Received03/07/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/25/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age74 YR
Patient SexFemale
-
-