• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDLINE EXCELSIOR; SYRINGE, 10ML SALINE IN 10ML

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDLINE EXCELSIOR; SYRINGE, 10ML SALINE IN 10ML Back to Search Results
Model Number EMZE010001
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/11/2023
Event Type  malfunction  
Event Description
It was reported that the pre-filled saline flush syringe would not fully thread onto a stopcock.
 
Manufacturer Narrative
It was reported that the pre-filled saline flush syringe would not fully thread onto a stopcock.To date, no information has been received to indicate that a user or a patient experienced a death, serious injury, medical intervention, follow-up care, or other adverse health impact associated with the reported problems/issues.A sample was returned for evaluation and damage to the luer collar threads was observed.The root cause was determined to be mishandling of the item.In an abundance of caution, and in response to an fda 483 issued for cfn (b)(4) on 22-jan-2024, this medwatch is being filed.If additional relevant information becomes available a supplemental medwatch will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
SYRINGE, 10ML SALINE IN 10ML
Manufacturer (Section D)
MEDLINE EXCELSIOR
1933 heck drive
neptune NJ 07753 4428
Manufacturer Contact
nigel vilches
three lakes drive
northfield, IL 60093-2753
MDR Report Key18865239
MDR Text Key337452978
Report Number2027791-2024-00016
Device Sequence Number1
Product Code NGT
UDI-Device Identifier20363807100019
UDI-Public20363807100019
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 03/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberEMZE010001
Device Lot Number3142652
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/16/2023
Initial Date Manufacturer Received 10/11/2023
Initial Date FDA Received03/08/2024
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-