MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC VASOVIEW HEMOPRO 2; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC VASOVIEW HEMOPRO 2; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number VH-4000
Device Problem Failure to Cut (2587)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/15/2024
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.
 
Event Description
The hospital reported that during an endoscopic vein harvesting procedure, vasoview hemopro 2 cautery failed to operate.A second evh kit was opened and the case finished without further incident.Delay to open new product.No harm was done to the patient.
 
Manufacturer Narrative
Trackwise#: (b)(4).The device was returned to the factory for evaluation on 02/23/2024.An investigation was conducted on 02/27/2024.A visual inspection was conducted.Signs of clinical use and evidence of blood and charred material were observed.The heater wire and clear silicone insulation of both the hot and cold jaw were observed to be intact with no visual defects.An electrical evaluation was conducted.A pre-cautery test was performed per the direction for use (dfu) with a reference cable, adapter, and reference power supply vh-3010 at level 3.0.The device passed the pre-cautery test; it produced visible steam and heat during ten (10) 3-second activations and shut off when the toggle was released.To evaluate the safety shut down system, a polyfuse activation test was performed 5 times over 10 minutes.The device shut off after the period of sustained activation and reactivated after 10-second cooling period with no incident each time.An activation and transection capability test was performed over four (4) repetitions using "max life test method stm2048073 rev aa.The device successfully transected tissue four (4) times.The jaws were gently cleaned of debris and char with a saline and gauze pad as indicated in the direction for use (cv000008979).A temperature and resistance test was conducted to evaluate the device function per hemopro 2 final test 90523436 rev w.The resistance value was measured at.686 ohms which is within specification.The device passed the temperature measurements test.The displayed temperature increased and turned green within the 2 second specified timeframe.The displayed temperature decreased once the toggle swivel was released.Based on the results of the evaluation, the reported failure "failure to cut" was not confirmed.The lot # 3000364115 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEW HEMOPRO 2
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key18866209
MDR Text Key337223217
Report Number2242352-2024-00217
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00607567700406
UDI-Public00607567700406
Combination Product (y/n)N
PMA/PMN Number
K101274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVH-4000
Device Catalogue NumberVH-4000
Device Lot Number3000364115
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2024
Initial Date Manufacturer Received 02/16/2024
Initial Date FDA Received03/08/2024
Supplement Dates Manufacturer Received03/28/2024
Supplement Dates FDA Received04/04/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/05/2024
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNKNOWN.
Patient Age71 YR
Patient SexMale
Patient Weight127 KG
-
-