Model Number WA53005A |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned, and the evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported, the telescope, 10 mm, 30°, hd, quick lock, autoclavable had a detached light guide part.There were no reports of patient harm.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the event was confirmed, and the device did not meet the standard specifications and function.The reported failure descriptions are most likely due to the effect of an excessive mechanical force caused by an impact or fall.No further information could be obtained.Olympus will continue to monitor field performance for this device.
|
|