MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; Stimulator, electrical, implanted, for essential tremor
|
Back to Search Results |
|
Model Number 37612 |
Device Problems
Charging Problem (2892); Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Continuation of d10: product id 37761, serial# (b)(6), product type recharger.Section d information references the main component of the system.Other relevant device(s) are: product id: 37761, serial/lot #: (b)(6).Analysis of the 37761 recharger (rtm) (b)(6 ) revealed broken connectors pins.This regulatory report is being submitted as part of a retrospective review as part of remediation plan 411.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the recharger was not charging up and had been plugged in to the desktop charger (dtc) for days, and that yesterday, the charge your recharger screen came up and they were not able to clear the screen or use the recharger.The caller stated the night before, they were able to use the recharger and get their ins from 25% to 50%.When asked, caller did not provide date event first started, and had mentioned that the desktop charger (dtc) connection in the recharger port was loose and had been coming disconnected for days.Caller confirmed green light of ac power was on.Caller stated there was no visible damage to the dtc, connector pin, or recharger port.The caller stated their ins was low at 25%.The issue was not resolved.No symptoms were reported.
|
|
Search Alerts/Recalls
|
|
|