ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; NO MATCH Back to Search Results
Model Number 1458Q/86
Device Problems High impedance (1291); Capturing Problem (2891); Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Injury  
Event Description
Related manufacturer reference number: 2017865-2024-34436.It was reported the patient presented in clinic for follow up.Upon interrogation, it was discovered the right ventricular and left ventricular leads exhibited high capture thresholds and decreased sensing.The leads were explanted and replaced.The patient was in stable condition.
 
Manufacturer Narrative
The reported events of unacceptable threshold, sensing anomaly, and high pacing impedance were not confirmed.A complete lead was returned in one piece for analysis.Electrical testing did not find any indication of conductor fractures or internal shorts.Visual and x-ray inspections of the lead did not find any anomalies.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18867803
MDR Text Key337244139
Report Number2017865-2024-34437
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734510219
UDI-Public05414734510219
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2022
Device Model Number1458Q/86
Device Lot NumberA000080788
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/18/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
QUADRA ASSURA; TENDRIL
Patient Outcome(s) Required Intervention;
Patient Age72 YR
Patient SexFemale
-
-