ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT DR; NO MATCH Back to Search Results
Model Number CDDRA500Q
Device Problems Failure to Interrogate (1332); Wireless Communication Problem (3283)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/15/2024
Event Type  malfunction  
Event Description
It was reported that a patient presented with loss of bluetooth telemetry noted on the patient's implantable cardioverter defibrillator.No intervention or adverse patient consequences were reported.
 
Event Description
New information received indicates the implantable cardioverter defibrillator was in bluetooth telemetry lockout.The device was re-interrogated and telemetry was restored.
 
Manufacturer Narrative
The reported event of ble unavailable was resolved by interrogating the device using a merlin programmer, this is indicative of a ble lockout which is per device design.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT DR
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18872407
MDR Text Key337563833
Report Number2017865-2024-34514
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032003
UDI-Public05415067032003
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 04/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2022
Device Model NumberCDDRA500Q
Device Lot NumberP000114838
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/20/2024
Initial Date FDA Received03/11/2024
Supplement Dates Manufacturer Received03/13/2024
04/11/2024
Supplement Dates FDA Received04/02/2024
04/11/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/07/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age72 YR
Patient SexMale
-
-