The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd drive pcb fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ccd drive pcb.In addition, our technician confirmed that the control body fluid damage, the u/d and the r/l pulley wires fluid damage, the electrical pin connector fluid damage, the lg connector fluid damage, the operation channel (primary) buckled, the shield cover corroded, the remote control buttons cut, the remote control switch defective, and the angle wire hard to move; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|